Immunoassay Analyzer Troubleshooting | Seamaty | Medical device manufacturers

release time:2021-08-20 16:02:39

In the previous article, Seamaty shared with you three common types of large immunoassay analyzers: fully automated chemiluminescence immunoassay analyzer, fully automated microparticle chemiluminescence immunoassay analyzer and fully automated electrochemiluminescence immunoassay analyzer. There are also maintenance methods for these immunoassay analyzers.


As we all know, these large immunoassay analyzers are complicated to operate and require specialized physician training to operate them. But the machine is inevitably malfunctioning. So if the equipment breaks down in, what can be done?

Chemiluminescent immunoassay analyzers are highly automated and generally have a self-diagnostic function. Once a fault occurs, the instrument can generally automatically detect, display error messages and accompanied by alarm sounds.

Common faults are mainly in the following areas.

1. The pressure gauge of the immunoassay analyzer indicates zero

Vacuum pressure test on the machine, you can hear the sound of the pump work, but the pressure gauge indication is zero.

First check the vacuum pipe connected to the waste liquid bottle, test the vacuum pressure and determine whether the fault is caused by air leakage or damage to the pressure gauge.

Then check the interface of each pipe for air leakage, check the four solenoid valves (in the vacuum pressure test, these four solenoid valves do not work, for the closed state), repair or replace the problematic pipes or solenoid valves in time.

2. Insufficient vacuum pressure of immunoassay analyzer

Vacuum pressure test on the analyzer, if the test result is normal, it is known that the vacuum sensor can not detect the vacuum pressure caused.
The machine's pressure test by two sensors to detect high and low pressure, the problem of the sensor to adjust or clean, and then test the vacuum pressure, the pressure is normal to adjust the sensor screw so that the high and low pressure indication in the specified range.

3. The analyzer luminous body error

a. Check the surface of the instrument luminous body found to have liquid seepage, the fault in three steps to check.
b. Check the waste liquid probe, related piping and cleaning pool for blockage and leakage.
c. Check the sample addition solenoid valve, drainage solenoid valve, solenoid valve, there is dirt will cause poor water intake or drainage.
d. Check whether there are air leaks in the alkali pump cleaning pipeline connected with the waste liquid probe pipeline and whether there are cracks in the alkali pump.

4. Wrong track of the instrument

The fault is caused by the misalignment of the specimen rack in the track and the track can not run. Because the track is very long, and closed not easy to dismantle, generally check the horizontal lift connected to the track first. If normal, then check the track, as long as the misaligned specimen rack is removed, the fault can be eliminated.

Seamaty Veterinary Immunoassay Analyzer SMT-680V

Immunoassay Analyzer SMT-680V

A. Compact and light weight

Veterinary Immunoassay Analyzer  SMT-680V Only 8Kg with small dimensions, space saving and fitting easily into your laboratory.

B. Minimum maintenance
SMT-680V Fluidics-free system for greater reliability, without carry over and less maintenance.

C. Easy to use
SMT-680V takes only several steps to finish the operation, increased efficiency and streamlines workflow.

Prev Return Next

Related Reading

13 Point-of-Care Testing Examples

2022-09-01

The development of POCT is driven by the rising demand for clinical testing. Although traditional testing methods can meet the requirements of clinical testing in terms of test specimen volume, automated operation, accuracy of results, performance stability, etc.. However, they are unable to meet the requirements in terms of portability, testing speed, and complexity of the testing process. In the family health management and primary care institutions testing, large medical testing equipment is difficult to spread. We took the hospital market as an example and sorted out the demand for POCT products in some hospital departments.

Animal Medicine: Coagulation Test And 4 Anticoagulants

2022-01-13

Coagulation refers to the ability of the body system to maintain the integrity of blood and blood vessels and is a complex interaction between the vessel wall, platelets and coagulation factors.

6 advantages of chemiluminescence immunoassay analyzer |Seamaty

2021-08-26

These new immunological analysis equipment has greatly increased the sensitivity and specificity of the analysis index, and also significantly improved the accuracy and efficiency of the test. In particular, the fully automated chemiluminescence immunoassay analyzer has become the most widely used and actively developed test equipment in clinical testing.