Guide to the use of coagulation reagent disc

release time:2021-04-22 09:30:03

The blood coagulation reagent with four indexes launched by Chengdu Seamaty Technology Co., Ltd. used with SMT-120V, an automatic biochemical analyzer, is suitable for preoperative examination of pets, which can help doctors to understand whether the animal's hemostatic function is defective. The preoperative examination can prevent intraoperative bleeding.


Guide to the use of coagulation reagent disc



Operation precautions:




1. Before blood collection, the reagent disc needs to be taken out of the refrigerated environment, placed in a room temperature environment between 10-40 degrees Celsius for 20 minutes, and then tested.

Warning: Do not use electrical appliances to increase the temperature of the reagent tray, such as using a hair dryer.



2. Centrifuge the empty sodium citrate anticoagulant tube



3. A purple scalp needle or a small needle syringe (1ml syringe) is required for blood collection



4. The blood collection process should be gentle and slow to avoid excessive force.


5. The time of blood collection should be controlled within one minute.


6. Remove the needle after blood collection, and add 0.5ml sample volume to the sodium citrate anticoagulation tube. Please pay attention to the scale on the anticoagulation tube.


7. Turn the anticoagulant tube upside down 8-10 times to mix, avoid violent shaking.


8. The test should be carried out immediately after sampling.


9. If it is not possible to test immediately, please store the sample in an environment of 2-8 degrees Celsius and test within two hours.


10. The sample types support whole blood or plasma


11. Before the test, if the sample has hemolysis or coagulation, the sample needs to be re-collected.


12. When adding samples to the reagent disc, the operation should be gentle to avoid bubbles.


Prev Return Next

Related Reading

Guide to the use of coagulation reagent disc

2021-04-22

The blood coagulation reagent with four indexes launched by Chengdu Seamaty Technology Co., Ltd. used with SMT-120V, an automatic biochemical analyzer, is suitable for preoperative examination of pets, which can help doctors to understand whether the animal's hemostatic function is defective. The preoperative examination can prevent intraoperative bleeding.

What should I do if I encounter an abnormal sample during pet biochemical testing?

2021-04-22

Compared to human medical testing, abnormal samples account for a larger number of blood samples that can be collected by veterinarians. When collecting blood, what should be done with abnormal samples such as lipemia, hemolysis, jaundice and coagulation?

Precautions for using veterinary biochemical analyzer SMT-120V

2021-04-21

Although the operation of the veterinary biochemical analyzer SMT-120V is simple, there are some precautions you need to know. The correct use of the biochemical analyzer can avoid abnormal results and provide correct data for clinical judgment.