SMT-30: Seamaty 3-part auto hematology analyzer
- 2 Reagents
- 21 Parameters
- 60 Samples/hour
- Low consumption of reagent and sample
Two modes of sampling: Whole blood, pre-diluted blood. Build-in thermal printer, external laser printer or inkjet printer. 4 USB, 1 LAN support protocol HL7 and compatible with LIS system. Accompanying quality control material
10.4 colorful touchscreen. Liquid Crystal Display. Resolution : 800600. 10°incline adapt to visual observation
Auto-cleaning of sample probe and tubes. Automatic fault processing function
Only need three reagents. Sample consumed ≤20uL
60 samples per hour
390mm(L)×285mm(W)×390mm(H) Weight≤16Kg
Parameter |
Precision |
Linear Range |
Contamination Rate |
WBC |
≤2.0% (3.5-15.0×10^9)/L |
(0-100.0×10^9)/L |
≤0.5% |
RBC |
≤1.5% (3.0-6.0×10^12)/L |
(0.10-8.00×10^12)/L |
≤0.5% |
HGB |
≤1.5% (100-180) g/L |
(0-250) g/L |
≤0.6% |
MCV |
≤1.0% (70-120)fL |
- |
- |
PLT |
≤4.0% (150-500×10^9)/L |
(0-1000×10^9)/L |
≤1.0% |
Power
Voltage: AC 100V~240V, 50/60Hz; Rated Power: 120VA
Principle
Electric impedance method and Colorimetry method
Parameters
WBC, LYM#, MID#, GRAN#, LYM%, MID%, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC
3 histograms
Reagents
Build-in LH Lyse, External Diluent, Probe Cleaner for maintenance
Flag System
RBC, WBC, PLT abnormal sample and support customized test groups
Storage
Up to 50,000 results including histograms, scattergrams and patient information.
Quality Control
3 Level QC, L-J graph, X-B
Calibration
Manual and Auto-calibration
Accessories
Standard: keyboard, waste liquid barrel, ground wire(3m); Option: mouse, external barcode scanner
Operational Environment
Working Temperature: 10℃~35℃; Relative Humidity: 20%~85%
Contact us for more details about this product, and we will get back to you within two working days.
Contact Us